Study on Micro Dried Bio-potential Electrodes Using Conductive Epoxy on Textile Fabrics
Doo Yeol Cha, Jung Mo Jung, Deok Su Kim, Hee Jun Yang, Kyo Sang Chol, Jong Myong Choi, Sung Pu Chang
J Electr Electron Mater 2013;26(5):367-372.   Published online May 1, 2013
In this paper, micro dried bio-potential electrodes are demonstrated for sEMG (surface ElectroMyoGraphic) signal measurement using conductive epoxy on the textile fabric. Micro dried hio-potential electrodes on the textile fabric substrate have several advantages over the conventional wetJdrv electrodes such as good feeling of wearing, possibility of extended-wearing due to the good ventilation. Also these electrodes on the textile fabric can easily apply to the curved skin surface. These electrodes are fabricated by the screen-printing process with the size of 1 mmxl0 mm and the resultant resistance of these electrodes have the average value of 0.4 Q. The conventional silver chloride electrode shows the average value of 0.3 . However, the electrode on the textile fabric are able to measure the sEMG signal without feeling of difference and this electrode shows the lower resistance of 1.03 &than conventional silver chloride electrode with 2.8 Q in the condition of the very sharp curve surface (the radius of curvature is 40 mm).